PAS 5500 Illumination: Slit Uniformity Operator:ASM Machine: Release:8.9.0 Date:05/03/2025 Time:02:22 OK: Test finished successfully Comment : MS_SLIT_UNIF Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 26.000 Y : 12.259 Steps X : 11 Y : 51 Number of Pulses : 100 Pulse Frequency [Hz] : 1000 Pulse Energy [mJ] : 10.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Conventional Numerical Aperture : 0.57 Sigma Outer : 0.750 Illumination mode name : conventional Version : 1.0.0 Uniformity Measurement Results Uniformity [%] : 3.68 X-Tilt [%/field] : -0.58 Symmetrical error [%] : 3.58 Logfile : LI/LISU/waf.1258 Overall Average of Ratio : 1.04 Standard Deviation : 0.02 Intensities Spot Sensor [mW/cm2] : 743.70 Standard Deviation : 17.79 Intensities Energy Sensor [mW/cm2] : 717.91 Standard Deviation : 0.08 Estimated Uniformity from measured data [%] If corrected for actual tilt : 3.82 If corrected with gradientfilter : 0.48 If corrected for tilt and with gradientfilter : 0.44 Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -0.097645 Position Y [mm] : 0.381051 Pointing X [mrad] : 0.196202 Pointing Y [mrad] : 0.036709 Laser beam-drift during measurement Position X [mm] : 0.062566 Position Y [mm] : -0.040794 Pointing X [mrad] : -0.002946 Pointing Y [mrad] : 0.001254 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 12.941150 | 1.036047 | | 10.352920 | 1.021671 | | 7.764690 | 1.000011 | | 5.176460 | 0.980138 | | 2.588230 | 0.967552 | | -0.000000 | 0.962454 | | -2.588230 | 0.972335 | | -5.176460 | 0.988955 | | -7.764690 | 1.008608 | | -10.352920 | 1.028817 | | -12.941150 | 1.033411 | +------------+--------------------+