PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:11/12/2024 Time:10:26 OK: Test finished successfully Comment : LS 0.51 0.36 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Quadrupole ID : 12-LOWSIGMA Numerical Aperture : 0.51 Sigma Outer : 0.360 Inner : 0.000 Illumination mode name : low sigma Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 0.84 X-Tilt [%/field] : 1.55 Symmetrical error [%] : 0.39 Logfile : LI/LISU/waf.68 Overall Average of Ratio : 1.01 Standard Deviation : 0.01 Intensities Spot Sensor [mW/cm2] : 791.76 Standard Deviation : 5.23 Intensities Energy Sensor [mW/cm2] : 782.35 Standard Deviation : 0.12 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.42 Unicom Filter id : Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -2.181241 Position Y [mm] : 0.441045 Pointing X [mrad] : 0.043714 Pointing Y [mrad] : 0.146190 Laser beam-drift during measurement Position X [mm] : -0.068817 Position Y [mm] : -0.056350 Pointing X [mrad] : -0.011965 Pointing Y [mrad] : -0.005421 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.011525 | | 5.551463 | 1.009259 | | 4.163597 | 1.004772 | | 2.775731 | 1.001166 | | 1.387866 | 0.998524 | | 0.000000 | 0.996820 | | -1.387866 | 0.994945 | | -2.775731 | 0.994906 | | -4.163597 | 0.994649 | | -5.551463 | 0.996207 | | -6.939328 | 0.997228 | +------------+--------------------+