PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:07/20/2026 Time:00:54 OK: Test finished successfully Comment : CON_0.60.42 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Conventional Numerical Aperture : 0.60 Sigma Outer : 0.420 Illumination mode name : conventional Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 0.67 X-Tilt [%/field] : 0.11 Symmetrical error [%] : 0.62 Logfile : LI/LISU/waf.493 Overall Average of Ratio : 1.02 Standard Deviation : 0.01 Intensities Spot Sensor [mW/cm2] : 676.35 Standard Deviation : 4.00 Intensities Energy Sensor [mW/cm2] : 666.28 Standard Deviation : 0.16 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.66 Unicom Filter id :-009 Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -3.051912 Position Y [mm] : 0.000311 Pointing X [mrad] : -0.031666 Pointing Y [mrad] : -0.012134 Laser beam-drift during measurement Position X [mm] : -0.095716 Position Y [mm] : 0.047666 Pointing X [mrad] : -0.034087 Pointing Y [mrad] : -0.005426 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.007847 | | 5.551463 | 1.003487 | | 4.163597 | 0.999931 | | 2.775731 | 0.997507 | | 1.387866 | 0.996207 | | 0.000000 | 0.995304 | | -1.387866 | 0.994398 | | -2.775731 | 0.996378 | | -4.163597 | 0.998398 | | -5.551463 | 1.003467 | | -6.939328 | 1.007076 | +------------+--------------------+