PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:11/18/2024 Time:03:15 OK: Test finished successfully Comment : CON_0.60.42 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Conventional Numerical Aperture : 0.60 Sigma Outer : 0.420 Illumination mode name : conventional Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 0.31 X-Tilt [%/field] : 0.56 Symmetrical error [%] : 0.10 Logfile : LI/LISU/waf.71 Overall Average of Ratio : 1.00 Standard Deviation : 0.00 Intensities Spot Sensor [mW/cm2] : 777.87 Standard Deviation : 1.95 Intensities Energy Sensor [mW/cm2] : 774.61 Standard Deviation : 0.15 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.17 Unicom Filter id : Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -2.139579 Position Y [mm] : 0.440910 Pointing X [mrad] : 0.033226 Pointing Y [mrad] : 0.140272 Laser beam-drift during measurement Position X [mm] : 0.069937 Position Y [mm] : -0.035447 Pointing X [mrad] : 0.023071 Pointing Y [mrad] : -0.004680 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.002736 | | 5.551463 | 1.003716 | | 4.163597 | 1.002372 | | 2.775731 | 1.001256 | | 1.387866 | 0.999756 | | 0.000000 | 0.998808 | | -1.387866 | 0.997593 | | -2.775731 | 0.997947 | | -4.163597 | 0.998596 | | -5.551463 | 0.998925 | | -6.939328 | 0.998296 | +------------+--------------------+