PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:07/19/2026 Time:22:24 OK: Test finished successfully Comment : LS 0.51 0.36 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Quadrupole ID : 12-LOWSIGMA Numerical Aperture : 0.51 Sigma Outer : 0.360 Inner : 0.000 Illumination mode name : low sigma Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 1.14 X-Tilt [%/field] : 1.01 Symmetrical error [%] : 0.81 Logfile : LI/LISU/waf.492 Overall Average of Ratio : 1.02 Standard Deviation : 0.01 Intensities Spot Sensor [mW/cm2] : 694.53 Standard Deviation : 5.75 Intensities Energy Sensor [mW/cm2] : 682.53 Standard Deviation : 0.15 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.86 Unicom Filter id :-009 Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -1.254914 Position Y [mm] : 0.033392 Pointing X [mrad] : 0.272870 Pointing Y [mrad] : -0.003173 Laser beam-drift during measurement Position X [mm] : 0.015753 Position Y [mm] : -0.043231 Pointing X [mrad] : 0.010364 Pointing Y [mrad] : -0.002515 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.014904 | | 5.551463 | 1.008019 | | 4.163597 | 1.001658 | | 2.775731 | 0.998303 | | 1.387866 | 0.996085 | | 0.000000 | 0.994407 | | -1.387866 | 0.992084 | | -2.775731 | 0.993811 | | -4.163597 | 0.994551 | | -5.551463 | 1.000819 | | -6.939328 | 1.005358 | +------------+--------------------+