PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:07/17/2026 Time:00:17 OK: Test finished successfully Comment : CON_0.60.42 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Conventional Numerical Aperture : 0.60 Sigma Outer : 0.420 Illumination mode name : conventional Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 0.54 X-Tilt [%/field] : 0.23 Symmetrical error [%] : 0.45 Logfile : LI/LISU/waf.491 Overall Average of Ratio : 1.02 Standard Deviation : 0.00 Intensities Spot Sensor [mW/cm2] : 687.93 Standard Deviation : 3.15 Intensities Energy Sensor [mW/cm2] : 675.54 Standard Deviation : 0.12 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.49 Unicom Filter id :-009 Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -2.899180 Position Y [mm] : -0.078662 Pointing X [mrad] : 0.008992 Pointing Y [mrad] : -0.027819 Laser beam-drift during measurement Position X [mm] : -0.122294 Position Y [mm] : -0.122750 Pointing X [mrad] : -0.022122 Pointing Y [mrad] : -0.009202 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.006083 | | 5.551463 | 1.003238 | | 4.163597 | 1.000608 | | 2.775731 | 0.998822 | | 1.387866 | 0.997547 | | 0.000000 | 0.996535 | | -1.387866 | 0.995364 | | -2.775731 | 0.996897 | | -4.163597 | 0.998470 | | -5.551463 | 1.002154 | | -6.939328 | 1.004283 | +------------+--------------------+