PAS 5500 Illumination: Slit Uniformity Operator:TEC Machine: Release:8.9.0 Date:11/15/2024 Time:02:21 OK: Test finished successfully Comment : CON_0.60.42 Measurement mode Scanning : No Start position : Left Side Meander direction : X-direction Measurement settings Fieldsize [mm] X : 14.000 Y : 11.271 Steps X : 11 Y : 33 Number of Pulses : 100 Pulse Frequency [Hz] : 4000 Pulse Energy [mJ] : 5.00 Miscellaneous Apply REMA Window : Yes Illumination Mode Illumination Mode : Conventional Numerical Aperture : 0.60 Sigma Outer : 0.420 Illumination mode name : conventional Version : 1.0.2 Uniformity Measurement Results Uniformity [%] : 0.46 X-Tilt [%/field] : 0.55 Symmetrical error [%] : 0.31 Logfile : LI/LISU/waf.69 Overall Average of Ratio : 1.01 Standard Deviation : 0.00 Intensities Spot Sensor [mW/cm2] : 773.31 Standard Deviation : 2.89 Intensities Energy Sensor [mW/cm2] : 767.19 Standard Deviation : 0.14 Estimated Uniformity from measured data [%] If corrected for actual tilt : 0.35 Unicom Filter id : Position [mm] : 32.40 Mode : Automatic Pre-expose Number of pre-expose pulses : 0 Initial Laser beam data Position X [mm] : -2.199615 Position Y [mm] : 0.373118 Pointing X [mrad] : 0.044871 Pointing Y [mrad] : 0.145254 Laser beam-drift during measurement Position X [mm] : 0.023235 Position Y [mm] : -0.010267 Pointing X [mrad] : 0.005826 Pointing Y [mrad] : -0.000460 Slit profile +------------+--------------------+ | Position X |Normalized Sum Ratio| | [mm] | | +============+====================+ | 6.939328 | 1.005449 | | 5.551463 | 1.004651 | | 4.163597 | 1.001873 | | 2.775731 | 1.000089 | | 1.387866 | 0.998295 | | 0.000000 | 0.997295 | | -1.387866 | 0.996178 | | -2.775731 | 0.996936 | | -4.163597 | 0.998098 | | -5.551463 | 1.000101 | | -6.939328 | 1.001036 | +------------+--------------------+